- P-ISSN 1225-0163
- E-ISSN 2288-8985
In this study, we improved the vitamin C test method and reviewed data on the adjustable range of chromatography conditions for quantification. First, we adjusted the mobile phase conditions such as solvent composition, salt concentration, pH and column temperature and in particular, it was confirmed through an improved test method that the peak derived from the buffer solution could be clearly separated from the target component, vitamin C by adjusting the pH. The retention time of vitamin C was partially changed by adjusting the column diameter, length and particle size but the number of theoretical plates indicated similar values and did not affect the separation and quantitative analysis of the target component. The flow rate according to the column specifications was derived from the equation proposed by the U.S. FDA (Food and Drug administration) and the Korean MFDS (Ministry of Food and Drug Safety), and evaluation of the applicability to vitamin complexes showed high selectivity for vitamin C even with altered stationary phase conditions and flow rates. In conclusion, vitamin C can be optimally separated and detected by changing the chromatographic method conditions and it was confirmed that the mobile and stationary phase conditions of liquid chromatography can be slightly adjusted in case the assay method uses an isocratic elution.