Feasibility of Using the Clamp Meter in Measuring X-Ray Tube Current

INTERNATIONAL JOURNAL OF CONTENTS / INTERNATIONAL JOURNAL OF CONTENTS, (P)1738-6764; (E)2093-7504
2013, v.9 no.1, pp.38-41
https://doi.org/10.5392/IJoC.2013.9.1.038
김성철 (가천대학교)

Abstract

The clamp meter maintains electric safety as a non-invasive method while measuring the absolute value of tube current with it has been recently developed for an X-ray high-tension cable. Especially this can show high accuracy at short X-ray exposure time. Considering such a condition, this study is to evaluate the feasibility of a clamp meter in measuring X-ray tube current by taking the measurements and comparing with those of the Dynalyzer III which has been considered as a standard measuring device. From measuring the tube current accuracy depending on changes in tube voltage and exposure time, the clamp meter showed higher accuracy rate which was -1.3~4.2% difference. Thus clamp meter can be used for clinical radiologists who are not familiar electric circuit to manage X-ray devices easily and correctly in the future.

keywords
X-ray tube current, Clamp meter, Non-invasive method, X-ray tube voltage