바로가기메뉴

본문 바로가기 주메뉴 바로가기

Reliability of Portable Spirometry Performed in the Korea National Health and Nutrition Examination Survey Compared to Conventional Spirometry

Tuberculosis & Respiratory Diseases / Tuberculosis & Respiratory Diseases,
2021, v.84 no.4, pp.274-281
https://doi.org/10.4046/trd.2021.0016
Hye Jung Park (Yonsei University)


Park Yong Bum (Hallym University Kangdong Sacred Heart Hospital)
  • Downloaded
  • Viewed

Abstract

Background: The Korea National Health and Nutrition Examination Survey (KNHANES) is a well-designed survey to collect national data, which many researchers have used for their studies. In KNHANES, although portable spirometry was used, its reliability has not been verified.Methods: We prospectively enrolled 58 participants from four Korean institutions. The participants were classified into normal pattern, obstructive pattern, and restrictive pattern groups according to their previous spirometry results. Lung function was estimated by conventional spirometry and portable spirometry, and the results were compared.Results: The intraclass correlation coefficients of forced vital capacity (FVC) (coefficient, 9.993; 95% confidence interval [CI], 0.988–0.996), forced expiratory volume in 1 second (FEV1) (coefficient, 0.997; 95% CI, 0.995–0.998), FEV1/FVC ratio (coefficient, 0.995; 95% CI, 0.992–0.997), and forced expiratory flow at 25–75% (FEF25–75%; coefficient, 0.991; 95% CI, 0.984–0.994) were excellent (all p<0.001). In the subgroup analysis, the results of the three parameters were similar in all groups. In the overall and subgroup analyses, Pearson’s correlation of all the parameters was also excellent in the total (coefficient, 0.986–0.994; p<0.001) and subgroup analyses (coefficient, 0.915–0.995; p<0.001). In the paired t-test, FVC, FEV1/FVC, and FEF25–75% estimated by the two instruments were statistically different. However, FEV1 was not significantly different.Conclusion: Lung function estimated by portable spirometry was well-correlated with that estimated by conventional spirometry. Although the values had minimal differences between them, we suggest that the spirometry results from the KNHANES are reliable.

keywords
Pulmonary Function Test, Reliability, Spirometry

Reference

1.

1. Kim Y, Park S, Kim NS, Lee BK. Inappropriate survey design analysis of the Korean National Health and Nutrition Examination Survey may produce biased results. J Prev Med Public Health 2013;46:96-104.

2.

2. Han H, Oh EY, Lee JH, Park JW, Park HJ. Effects of particulate matter 10 inhalation on lung tissue RNA expression in a murine model. Tuberc Respir Dis 2021;84:55-66.

3.

3. Kim SY, Kim E, Kim WJ. Health effects of ozone on respiratory diseases. Tuberc Respir Dis 2020;83(Suppl 1):S6-11.

4.

4. Jirjees FJ, Dallal Bashi YH, Al-Obaidi HJ. COVID-19 death and BCG vaccination programs worldwide. Tuberc Respir Dis 2021;84:13-21.

5.

5. Kwon DS, Kim TH, Byun MK, Kim HJ, Lee HS, Park HJ, et al. Positive effects of the national cigarette price increase policy on smoking cessation in South Korea. Tuberc Respir Dis 2020;83:71-80.

6.

6. Park YS, Park S, Lee CH. The attributable risk of smoking on all-cause mortality in Korean: a study using KNHANES IV-VI (2007-2015) with mortality data. Tuberc Respir Dis 2020;83:268-75.

7.

7. Park HJ, Byun MK, Kim HJ, Kim JY, Kim YI, Yoo KH, et al. Dietary vitamin C intake protects against COPD: the Korea National Health and Nutrition Examination Survey in 2012. Int J Chron Obstruct Pulmon Dis 2016;11:2721-8.

8.

8. Yoo KH, Kim YS, Sheen SS, Park JH, Hwang YI, Kim SH, et al. Prevalence of chronic obstructive pulmonary disease in Korea:the fourth Korean National Health and Nutrition Examination Survey, 2008. Respirology 2011;16:659-65.

9.

9. Chung SJ, Kim HI, Yang B, Kim T, Sim YS, Kang HK, et al. Impact of the severity of restrictive spirometric pattern on nutrition, physical activity, and quality of life: results from a nationally representative database. Sci Rep 2020;10:19672.

10.

10. Izbicki G, Abboud S, Jordan P, Perruchoud AP, Bolliger CT. A comparison of a new transtelephonic portable spirometer with a laboratory spirometer. Eur Respir J 1999;14:209-13.

11.

11. Liistro G, Vanwelde C, Vincken W, Vandevoorde J, Verleden G, Buffels J, et al. Technical and functional assessment of 10office spirometers: a multicenter comparative study. Chest 2006;130:657-65.

12.

12. Park YB, Rhee CK, Yoon HK, Oh YM, Lim SY, Lee JH, et al. Revised (2018) COPD clinical practice guideline of the Korean Academy of Tuberculosis and Respiratory Disease: a summary. Tuberc Respir Dis 2018;81:261-73.

13.

13. Lee SH, Yeo Y, Kim TH, Lee HL, Lee JH, Park YB, et al. Korean guidelines for diagnosis and management of interstitial lung diseases: part 2. idiopathic pulmonary fibrosis. Tuberc Respir Dis 2019;82:102-17.

14.

14. Miller MR, Hankinson J, Brusasco V, Burgos F, Casaburi R, Coates A, et al. Standardisation of spirometry. Eur Respir J 2005;26:319-38.

15.

15. Aaron SD, Tan WC, Bourbeau J, Sin DD, Loves RH, MacNeil J, et al. Diagnostic instability and reversals of chronic obstructive pulmonary disease diagnosis in individuals with mild to moderate airflow obstruction. Am J Respir Crit Care Med 2017;196:306-14.

16.

16. Schermer TR, Robberts B, Crockett AJ, Thoonen BP, Lucas A, Grootens J, et al. Should the diagnosis of COPD be based on a single spirometry test? NPJ Prim Care Respir Med 2016;26:16059.

17.

17. Singh D, Agusti A, Anzueto A, Barnes PJ, Bourbeau J, Celli BR, et al. Global strategy for the diagnosis, management, and prevention of chronic obstructive lung disease: the GOLD science committee report 2019. Eur Respir J 2019;53:1900164.

18.

18. Ramos Hernandez C, Nunez Fernandez M, Pallares Sanmartin A, Mouronte Roibas C, Cerdeira Dominguez L, Botana Rial MI, et al. Validation of the portable air-smart spirometer. PLoS One 2018;13:e0192789.

19.

19. Kupczyk M, Hofman A, Koltowski L, Kuna P, Lukaszyk M, Buczylko K, et al. Home self-monitoring in patients with asthma using a mobile spirometry system. J Asthma 2021;58:505-11.

20.

20. Yang SY, Kim YH, Byun MK, Kim HJ, Ahn CM, Kim SH, et al. Repeated measurement of fractional exhaled nitric oxide is not essential for asthma screening. J Investig Allergol Clin Immunol 2018;28:98-105.

Tuberculosis & Respiratory Diseases