Validation of the Analytical Procedure for Quantitative Determination of Four Trace Metals (As, Cd, Pb, and Hg) in Fish Lipids Using Inductively Coupled Plasma-Mass Spectrometry (ICP-MS)

Mass Spectrometry Letters / Mass Spectrometry Letters, (P)2233-4203; (E)2093-8950
2024, v.15 no.3, pp.149-157
https://doi.org/10.5478/MSL.2024.15.3.149
Jayakody Kasun S. (Rajarata University of Sri Lanka)
Edirisinghe Ranjith K. B. (Rajarata University of Sri Lanka)
Senevirathne Suchithra A. (Rajarata University of Sri Lanka)
Senarathna Lalith (Rajarata University of Sri Lanka)
  • Downloaded
  • Viewed

Abstract

The objective of the present study was to validate the analytical procedure for the quantitative determination of four trace metals (As, Cd, Pb, and Hg) in extracted fish lipids using Inductively Coupled Plasma-Mass Spectrometry, ICP-MS. The extracted lipids using Bligh and Dyer method were digested by means of microwave-assisted acid digestion and introduced into an optimized ICP-MS instrument. The validation of the analytical method was carried out in accordance with the international standards and guidelines outlined in the European Pharmacopeia (2022), which included specificity, selectivity, linearity, limit of detection, limit of quantification, precision, and accuracy. The linearity ranges of the calibration curves were R 2 > 0.999, while the relative standard deviation (%RSD) for precision was within 5%. All targeted trace metals have shown mean recoveries between 88.0%–114.9%. The obtained LOD and LOQ values for this analytical protocol indicated the ability to detect and quan- tify of As, Cd, Pb, and Hg at trace levels. The overall validation confirms the described analytical method was appropriate for routine analyses of As, Cd, Pb, and Hg in fish lipids.

keywords
Method validation, Fish lipids, Trace metals, ICP-MS


Submission Date
2024-02-22
Revised Date
2024-08-21
Accepted Date
2024-09-02