논문 상세

Home > 논문 상세
  • P-ISSN 2233-4203
  • E-ISSN 2093-8950

Isotopic Analysis of NUSIMEP-6 Uranium Particles using SEM-TIMS

Mass Spectrometry Letters / Mass Spectrometry Letters, (P)2233-4203; (E)2093-8950
2013, v.4 no.3, pp.51-54
https://doi.org/10.5478/MSL.2013.4.3.51
Jong-Ho Park (Korea Atomic Energy Research Institute)
Sujin Park (Korea Atomic Energy Research Institute)
Kyuseok Song (Korea Atomic Energy Research Institute)
  • 다운로드 수
  • 조회수

Abstract

Isotopic analysis using thermal ionization mass spectrometry coupled with scanning electron microscopy (SEMTIMS)was performed to determine the isotopic ratios of uranium contained in micro-particles in the 6th Nuclear SignaturesInterlaboratory Measurement Evaluation Programme (NUSIMEP-6) sample. Elemental analysis by energy dispersive X-rayspectroscopy (EDS) was conducted on uranium-bearing mirco-particles, which were transferred to rhenium filaments for TIMSloading using a micromanipulation system in a SEM. A multi-ion-counter system was utilized to detect the ion signals of thefour isotopes of uranium simultaneously. The isotope ratios of uranium corrected by bracketing using a reference materialshowed excellent agreement with the certified values. The measurement accuracy for n(234U)/n(238U) and (b) n(235U)/n(238U) was10% and 1%, respectively, which met the requirements for qualification for the NetWork of Analytical Laboratories (NWAL).

keywords
Uranium, SEM-TIMS, Particle Analysis, Isotopic Analysis, Safeguards


참고문헌

1

Donohue, D. L. (1998). . J. Alloy Compd, 271-273, 11-.

2

Heumann, K. G. (1995). . Analyst, 120, 1291-.

3

Esaka, F. (2009). . Talanta, 78, 290-.

4

Fumitaka Esaka. (2011). Feasibility Study of Isotope Ratio Analysis of Individual Uranium-Plutonium Mixed Oxide Particles with SIMS and ICP-MS. Mass Spectrometry Letters, 2(4), 80-83.

5

Lee, C. G. (2006). . Jpn. J. Appl. Phys, 45, 294-.

6

Esaka, K. T. (2004). . Jpn, J. Appl. Phys, 43, L915-L9164.

7

Jong Ho Park. (2010). A Technique to Minimize Impurity Signal from Blank Rhenium Filaments for Highly Accurate TIMS Measurements of Uranium in Ultra-Trace Levels. Mass Spectrometry Letters, 1(1), 17-20.

8

Kyuseok Song. (2011). A Correction Method for the Peak Tailing Backgrounds for Accurate Isotope Ratio Measurements of Uranium in Ultra Trace Levels using Thermal Ionization Mass Spectrometry. Bulletin of the Korean Chemical Society, 32(12), 4327-4331.

9

Park, J.-H. (2013). . Asian J. Chem, 25, 7061-.

10

Suzuki, D. (2010). . Int. J. Mass. Spectrom, 294, 23-.

11

Bürger, S. (2010). . Int. J. Mass Spectrom, 294, 65-.

투고일Submission Date
2013-09-11
수정일Revised Date
2013-09-24
게재확정일Accepted Date
2013-09-24
상단으로 이동

Mass Spectrometry Letters