논문 상세

Home > 논문 상세
  • P-ISSN 2233-4203
  • E-ISSN 2093-8950

Fast Screening of Harmful Disinfectants in Household Products via Low-Temperature Plasma Ionization-Mass Spectrometry

Mass Spectrometry Letters / Mass Spectrometry Letters, (P)2233-4203; (E)2093-8950
2017, v.8 no.2, pp.44-47
https://doi.org/10.5478/MSL.2017.8.2.44
Hyoung Jun Lee (Korea Research Institute of Standard and Science, Chungnam National University)
,Gi Ryang Kweon (Chungnam National University)
Yong-Hyeon Yim (Korea Research Institute of Standard and Science)
  • 다운로드 수
  • 조회수

Abstract

Isothiazolinone derivatives are widely used in consumer products as disinfectants or preservatives, but there are growing concerns about their impact on human health. Therefore, rapid screening of these biocides is very important for proper control and regulation of potentially hazardous substances. To this end, low-temperature plasma (LTP) ionization mass spectrometry (MS) was investigated to demonstrate its potential for direct and selective analysis of isothiazolinones from sprayed aerosol samples. Benzisothiazolinone (BIT) was clearly identified from a commercial fabric deodorant using LTP ionization MS and MS/MS. LTP allowed selective ionization of BIT directly from the simply sprayed aerosol sample and illustrated its potential for fast screening without sample pre-treatments. Selective nature of LTP ionization, on the other hands, implicates use of LTP ionization MS as a general screening method for specific groups of hazardous chemicals in commercial products.

keywords
low-temperature plasma ionization, benzisothiazolinone, fabric deodorant, fast screening


참고문헌

1

Garcia-Hidalgo, E. (2017). . Contact Dermatitis, 76, 96-.

2

Kim, K. W. (2014). . Am. J. Respir. Crit. Care Med, 189, 48-.

3

Kwon, J.-T. (2013). . Mol. Cell Toxicol, 9, 51-.

4

Chang, M.-H. (2012). . Environ. Health Toxicol, 27, e2012003-.

5

Aerts, O. (2017). . Eur. J. Dermatol, 27, 115-.

6

Lin, Q. -B. (2010). . Food Addit. Contam, 27, 1775-.

7

Bester, K. (2010). . J. Chromatogr. A, 1217, 5204-.

8

임용현. (2015). Peltier Heating-Assisted Low Temperature Plasma Ionization for Ambient Mass Spectrometry. Mass Spectrometry Letters, 6(3), 71-74. http://dx.doi.org/10.5478/MSL.2015.6.3.71.

9

Harper, J. D. (2008). . Anal. Chem, 80, 9097-.

10

Takats, Z. (2004). . Science, 306, 471-.

11

Cody, R. B. (2005). . Anal. Chem, 77, 2297-.

투고일Submission Date
2017-06-17
수정일Revised Date
2017-06-22
게재확정일Accepted Date
2017-06-23
상단으로 이동

Mass Spectrometry Letters