Dynamic MRM Measurements of Multi-Biomarker Proteins by Triple- Quadrupole Mass Spectrometry with Nanoflow HPLC-Microfluidics Chip

Mass Spectrometry Letters / Mass Spectrometry Letters, (P)2233-4203; (E)2093-8950
2010, v.1 no.1, pp.21-24
https://doi.org/10.5478/MSL.2010.1.1.021
Ji Eun Sun (Korea Basic Science Institute,Hannam University)
Cheon Mi Hee (Korea Basic Science Institute,Yonsei University)
Lee Ju Yeon (Korea Basic Science Institute,)
Yoo Jong Shin (Korea Basic Science Institute, Chungnam National University)
Jung Hyun-Jin (Agilent technologies Korea Ltd.)
Kim Jin Young (Korea Basic Science Institute)
  • Downloaded
  • Viewed

Abstract

The development of clinical biomarkers involves discovery, verification, and validation. Recently, multiple reactionmonitoring (MRM) coupled with stable isotope dilution mass spectrometry (IDMS) has shown considerable promise for the directquantification of proteins in clinical samples. In particular, multiple biomarkers have been tracked in a single experiment usingMRM-based MS approaches combined with liquid chromatography. We report here a highly reproducible, quantitative, anddynamic MRM system for validating multi-biomarker proteins using Nanoflow HPLC-Microfluidics Chip/Triple-QuadrupoleMS. In this system, transitions were acquired only during the retention window of each eluting peptide. Transitions with the highestMRM-MS intensities for the five target peptides from colon cancer biomarker candidates were automatically selected usingOptimizer software. Relative to the corresponding non-dynamic system, the dynamic MRM provided significantly improvedcoefficients of variation in experiments with large numbers of transitions. Linear responses were obtained with concentrationsranging from fmol to pmol for five target peptides.

keywords
Dynamic MRM, HPLC-Microfluidics Chip, Protein quantitation, Reproducibility, Biomarker validation


Reference

1

Ong, S. E. (2005). . Nat. Chem. Biol, 1, 252-.

2

Anderson, L. (2006). . Mol. Cell. Proteomics, 5, 573-.

3

Fortin, T. (2009). . Anal. Chem, 81, 9343-.

4

Nicol, G. R. (2008). . Mol. Cell. Proteomics, 7, 1974-.

5

Sherwood, C. A. (2009). . J. Proteome Res, 8, 3746-.

6

Kuzyk, M. A. (2009). . Mol. Cell. Proteomics, 8, 1860-.

7

Stahl-Zeng, J. (2007). . Mol. Cell. Proteomics, 6, 1809-.

8

Tang, N. (2009). (-). ASMS Conference on Mass Spectrometry.

9

Alley, W. R. (2010). . Anal. Chem, 82, 5095-.

10

Faure, K. (2010). . Electrophoresis, 31, 2499-.

Submission Date
2010-11-14
Revised Date
2010-11-27
Accepted Date
2010-11-27
상단으로 이동

Mass Spectrometry Letters