Unusual Applications of Kendrick Plots: Recalibration and Tolerance

Mass Spectrometry Letters / Mass Spectrometry Letters, (P)2233-4203; (E)2093-8950
2023, v.14 no.4, pp.173-177
https://doi.org/10.5478/MSL.2023.14.4.173
Thierry N. J. Fouquet (Surgical Analytical & Materials Science (SAMS))
Orlando Cabarcos (Surgical Analytical & Materials Science (SAMS))
  • 다운로드 수
  • 조회수

Abstract

Kendrick plots offer an alternative visualization of mass spectral data which reveals ion series and patterning by turn- ing a mass spectrum into a map, plotting the fractional mass (wrongly called mass defect) as a function of mass-to-charge ratios and ion abundances. Although routinely used for polymer mass spectrometry, two unreported applications of these Kendrick plots are proposed using the program “kendo2”: the graphical recalibration of a mass spectrum via the simulation of a theoretical fractional mass and a multi-segment fit; and the rapid evaluation of scan-to-scan variation of accurate mass measurements used as tolerances for the blank subtraction of UPLC-MS data files. Both applications are compatible with any type of high-resolution MS data including LC/GC-MS(/MS).

keywords
Kendrick analysis, polymer mass spectrometry, LC-MS, calibration, blank subtraction, data processing