The focus of this paper is to present techniques to overcome certain difficulties in quantitative analysis with a time- of-flight mass spectrometer (TOF-MS). The method is based on conventional solid-phase extraction, followed by reversed- phase ultra high performance liquid chromatography of the extract, and mass spectrometric analysis. The target compounds included atenolol, atrazine, caffeine, carbamazepine, diclofenac, estrone, ibuprofen, naproxen, simazine, sucralose, sulfamethox- azole, and triclosan. The matrix effects caused by high concentrations of organic compounds in wastewater are especially signif- icant in electrospray ionization mass spectroscopy. Internal-standard calibration with isotopically labeled standards corrects the results for many matrix effects, but some peculiarities were observed. The problems encountered in quantitation of carbamaze- pine and triclosan, due to nonlinear calibration were solved by changing the internal standard and using a narrower mass win- dow. With simazine, the use of a quadratic calibration curve was the best solution.
Ferrer, I.. (2003). . Trends Anal. Chem., 22, 750-. http://dx.doi.org/10.1016/S0165-9936(03)01013-6.
Petrovic, M.. (2003). . Trends Anal. Chem., 22, 685-. http://dx.doi.org/10.1016/S0165-9936(03)01105-1.
Lindholm, P. C.. (2014). . BioResources, 9, 3688-.
Rodriguez-Chueca, J.. (2018). . J. Hazardous Mater., , -. http://dx.doi.org/10.1016/j.jhazmat.2018.04.044.
Gomez, M. J.. (2010). . J. Chromatogr. A, 1217, 7038-. http://dx.doi.org/10.1016/j.chroma.2010.08.070.
Meng, C. -K.. Agilent Application Note 5989-5319EN.
Al-Qaim, F. F.. (2014). . J. Braz. Chem. Soc., 25, 271-.