- P-ISSN 2233-4203
- E-ISSN 2093-8950
The utilization of methyl 2-aminobenzoate as a biocide and pesticide has raised concerns regarding its potential tox- icity. To assess its safety, it is crucial to determine its quantity and related toxicity using reference materials (RMs) or certified reference materials (CRMs). As an RM and CRM containing methyl 2-aminobenzoate within a biocidal product matrix is cur- rently unavailable, this study aimed to produce a high-quality RM containing methyl 2-aminobenzoate, ensuring its homogene- ity and stability, following the ISO Guide 35 and ISO 17034. The study determined that the produced RM exhibited homogeneity, as indicated by a calculated F-value (1.91) smaller than the critical F-value (3.02). In the assessment of isochro- nous short-term stability, the slope of the linear regression for the RM showed no statistically significant difference from zero when stored at temperatures of 4, 18, and 60 °C for 4 weeks. Regarding classical long-term stability, the RM demonstrated sustained stability over the course of one year when stored at 4 °C. This study has successfully developed an RM for monitoring methyl 2-aminobenzoate in biocides and pesticides. Its quality underwent rigorous evaluation, confirming both homogeneity and stability.