Determination of Cd and Pb in Human Blood by Isotope Dilution Inductively Coupled Plasma Mass Spectrometry : International Comparison

Analytical Science and Technology / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
1996, v.9 no.2, pp.145-160
Park, C.J.
Suh, J.K.
Lee, S.H.
  • Downloaded
  • Viewed

Abstract

Inorganic analytical laboratory of Korea Research Institute of Standards and Science participated in an interlaboratory comparison program operated by Quebec Toxicology Centre of Canada in 1994 and again in 1995. The objective of this program is to enable participating laboratories to assess reproducibility and accuracy of their analytical results for trace toxic elements in human biological fluids. This laboratory determined Cd and Pb concentrations in 3 levels of human blood samples by isotope dilution inductively coupled plasma mass spectrometry. 0.5mL of blood sample is added to the digestion bomb together with 2mL of nitric acid and enriched spike isotopes and then decomposed in the microwave digestion system. The decomposed sample is diluted to 10mL and nebulized into ICP-MS. The Cd and Pb values reported by all participating laboratories are presented and compared. The values reported by this laboratory are within the acceptable range of target values.

keywords
Interlaboratory comparision, Cd and Pb, Human blood, ICP-MS