- P-ISSN 1225-0163
- E-ISSN 2288-8985
An accurate analysis method for the determination of lead in whole blood by ICP/MS was developed. Whole blood samples were decomposed in microwave digestion system without any contamination and loss of lead. The 96 samples were analyzed by ICP/MS using mass<TEX>$^{208}$</TEX> isotope of lead. Lead concentrations of human whole blood were ranged of <TEX>$2.50{\sim}22.8{\mu}g/dL$</TEX>. The accuracy of this analysis method was verified by analyzing of NIST SRM 955a series(lead in blood).