- P-ISSN 1225-0163
- E-ISSN 2288-8985
A single use, screen-printed sensor for the measurement of liquid phase ethanol was developed and its electrochemical performance was investigated. Disposable type edthanol sensor was fabricated by serially screen printing the carbon paste, silverd pasted and insulator inlon a polyester substrate to pattern working and reference electrode sites and electrical contact. Alcohol dehydrogenase(ADH) or alcohol oxidase(AOD) together with appropriate electron transfer mediators was immobilized on the working electrode. To improve the sensitivity and reproducibility of carbon paste electrode, some pretreatment procedures were applied and their resultant electrochemical performance was examined. The disposable type electrochemical ethanol sensor developed in this study conveniently determines the ethanol in liquid samples such as blood and in fermentation process.