흡착벗김 전압전류법에 의한 시안화이온의 정량

Determination of Cyanide Ion by Adsorptive Stripping Voltammetry

분석과학 / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
2001, v.14 no.6, pp.540-544
차기원
고수진
김재균
  • 다운로드 수
  • 조회수

Abstract

Determination of cyanide ion has been studied by adsorptive stripping voltammetry using hanging mercury electrode. Cyanide ion complexed with copper ion is adsorpbed on the electrode and oxidised at the positive potential scan. Optimal conditions of CN determination were found to be ; supporting electrolyte solution ; 0.1 M NaCl of ammonium buffer at pH 10, accumulation potential; -800 mV vs Ag/AgCl, accumulation time ; 300 s, scan rate ; 50mV/s. The linear concentration of cyanide ion was observed in the range <TEX>$1{\times}10^{-8}$</TEX>, <TEX>$1{\times}10^{-7}M$</TEX>. The detection limit(n/s

keywords
cyanide ion, adsorptive stripping voltammetry, <TEX>$Cu^{2+}$</TEX>