Article Detail

Home > Article Detail
  • P-ISSN 1225-0163
  • E-ISSN 2288-8985

Deermination of phenobarbital in Intact Phenobabital Tablets using NIRS

Analytical Science and Technology / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
2002, v.15 no.2, pp.102-107







  • Downloaded
  • Viewed

Abstract

This paper describes a rapid determination of phenobarbital in intact phenobarbital tablets using partial least squares regression(PLSR) method of transmittance spectrum of near infrared (NIR) compared with the analytical data of liquid chromatograpy. The linearity, concentration range and precision of this analytical method are studied. The correlation coefficient of the calibration curve is 0.9983 and the standard error of calibration(SEC) is 0.14 %. Intra-day precision and Inter-day precision obtained in this method are CV

keywords
near infrared spectroscopy, partial least square regression, phenobarbital