Comparisons of sample preparation and measurement (inductively coupled plasma mass spectrometry and graphite furnace atomic absorption spectrometry) in the determination of bone lead.

Analytical Science and Technology / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
2003, v.16 no.2, pp.152-158




  • Downloaded
  • Viewed

Abstract

This study was conducted to evaluate two sample digestion procedures and instrumental determination parameters for analysis of lead in bone. Amputated human legs were treated by acid digestion or microwave dissolution prior to spectrometric analysis. Inductively coupled plasma mass spectrometry (ICP-MS) and graphite furnace atomic absorption spectrometry (GF-AAS) were used for determining bone lead levels. Recovery efficiencies using standard reference material from acid digestion measured by ICP-MS were in good agreement with those of the certified value, but in cases of acid digestion by GF-AAS and microwave digestion by both two methods, recovery underestimated and overestimated, respectively. For the bone samples, the lead concentrations obtained by ICP-MS after acid digestionwere in good agreement with those by GF-AAS (correlation coefficient

keywords
bone lead, acid digestion, microwave oven, SRM, ICP-MS, GF-AAS