- P-ISSN 1225-0163
- E-ISSN 2288-8985
A high performance liquid chromatography gradient elution method with fluorescence detection for the determination of biotin in pure form and pharmaceutical preparations has been developed. BrMDMC gives intense fluorescence and the fluorescence was monitored with excitation at 360 nm and emission at 410 nm. The calibration curve for biotin shows good linearity over the range of 5 ~ 400 ng with correlation coefficient of 0.999. The detection limit of biotin was 2 ng and the result of recovery was 98.75% with relative standard deviation of 1.1%.