Article Detail

Home > Article Detail
  • P-ISSN 1225-0163
  • E-ISSN 2288-8985

Determination of osmium using sulfurous acid as reductant by ICP-AES

Analytical Science and Technology / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
2007, v.20 no.3, pp.251-254


  • Downloaded
  • Viewed

Abstract

In this study, a convenient, fast and accurate inductively coupled plasma atomic emission spectrometry (ICP-AES) method has been optimized for the determination of osmium in aqueous solutions. The method makes use of the reaction of sulfurous acid with osmium to quantitative conversion of volatile Os(VIII) to non-volatile Os(IV) in the pH range 2-10. The response was found to be stabilized immediately after sulfurous acid reacted with osmium. The precision was calculated to be 0.5-4.5 % (RSD) under various ICP-AES conditions. The detection limit was 2.5-57.7 ng/g based on <TEX>$3{\sigma}$</TEX> of the blank response (n

keywords
osmium analysis, sulfurous acid, ICP-AES