Article Detail

Home > Article Detail
  • P-ISSN 1225-0163
  • E-ISSN 2288-8985

Article Contents

    Simultaneous determination of low molecular weight amines and quaternary ammonium ions by IC/ESI-MS

    Analytical Science and Technology / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
    2007, v.20 no.3, pp.255-260



    • Downloaded
    • Viewed

    Abstract

    A new method for the simultaneous determination of low molecular weight amines and quaternary ammonium ions based on the separation by IC with a suppressor and the detection by MS with ESI has been developed. The method has been applied to the analysis of a mixture containing tetramethylammonium ion, tetraethylammonium ion, tetrapropylammonium ion, triethanolamine, trimethylamine and triethylamine. The constituents were separated by isocratic elution using an IonPac CS17 column, a cation-exchange column, and detected by conductivity and mass spectrometry. The newly developed method for the six components demonstrated that the repeatability in terms of relative standard deviation for three measurements was in the range of 0.1-0.5 %. The detection limits were between 0.2 and <TEX>$0.9{\mu}g/mL$</TEX> by the IC/ESI-MS.

    keywords
    IC/ESI-MS, low molecular weight amine, quaternary ammonium ion


    Reference

    1

    S. D. Nelson, (1985) “Bioactivation of Foreign Compounds”, Academic Press