- P-ISSN 1225-0163
- E-ISSN 2288-8985
The purpose on this study was to develop a new improved chromatographic method for determination of trace phenols from environmental waste water. The research was carried out with selected 8 phenols, and solid-phase extraction was employed as sample pretreatment method. The coupling of XAD-4 and Dowex <TEX>$1{\times}8$</TEX> resin as preconcentration column increased the selectivities toward interferences coexisted in matrix. Automation was accomplished with on-line process of pretreatment and HPLC system. After elution of sample through XAD-4 column, phenols were adsorbed by dispersion force, then displaced from it by ACN basified, simultaneously and selectively readsorbed via anion exchange on Dowex <TEX>$1{\times}8$</TEX>. Dowex <TEX>$1{\times}8$</TEX> column was washed by water. Phenols readsorbed were removed from Dowex <TEX>$1{\times}8$</TEX> column by a minimum volumn of methanol containing HCl. Each pretreatment step was connected by switching valves and the eluate was directly on-line injected to obtain fast and reliable results into the HPLC. Recovery of phenols was greater than 90%. To examine utility of this method, analysis of phenols from laboratory waste water sample which was added some organic pollutants to find with phenols on environmental waste water were also accomplished without their interference effects.