Evaluation of Quantitative Results of PIXE Analysis

Analytical Science and Technology / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
1997, v.10 no.5, pp.386-394
Park, Jeong-Ho
Choi, Kum-Chan
  • Downloaded
  • Viewed

Abstract

The standard samples for the calibration of PIXE analysis were prepared using the single element standard samples and sulfur and chlorine compounds. The diluted standard solutions were dropped on the non-hole nuclepore film and dried. Standard sample of six elements of Ca, Cr, Co, Cu, Se and Sr had close agreement between the measurement and theoretical values. According to the theoretical study about the effects of thick samples on the accuracy of the PIXE analysis, the reduction of X-ray yields in the thick sample can not be neglected for the sample thicker than <TEX>$100{\mu}g/cm^2$</TEX>.

keywords
PIXE analysis, Standard sample, Calibration of PIXE, Effects of thick sample