Quantitative Analysis of Acylcarnitines and Amino Acids Using ESI-MS/MS

Analytical Science and Technology / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
2001, v.14 no.5, pp.384-391
Kim, Hohyun
Han, Sang-Beom
Yoon, Hye-Ran
  • Downloaded
  • Viewed

Abstract

In this study, a new quantitative analytical method has been developed for the rapid determination of acylcarnitines and amino acids in human blood using electrospray ionization / tandem mass spectrometry (ESI-MS/MS). Acylarmitines and amino acids were analyzed by tandem mass spectrometry after conversion to their butylesters through treatment with 3N butanolic hydrogen chloride. Acylcrnitines were analyzed using precursor 85 ion scan and alanine, valine, leucine/isoleucine, methionine, phenylalanine, tyrosine, aspartic acid and glutamic acid were analyzed using neutral loss 102 scan, ornitine and citrulline were analyzed neutral loss 119 scan, glycine was analyzed using neutral loss 56 scan, arginine was analyzed using neutral loss 161 scan and argininosuccinic acid was analyzed product ion 459 scan. This method reduced sample preparation time compared to that with conventional amino acid analyzer and liquid chromatography, with high sensitivity and good reproducibility.

keywords
electrospray ionization, tandem mass spectrometry, acylcarnitine, amino acid