• P-ISSN1225-0163
  • E-ISSN2288-8985
  • SCOPUS, ESCI, KCI

Article Detail

Home > Article Detail
  • P-ISSN 1225-0163
  • E-ISSN 2288-8985

Matrix effect of Ti and Zr-2.5Nb sample for hydrogen analysis by Inert Gas Fusion-Thermal Conductivity Deterction(IGE-TCD) Method

Analytical Science and Technology / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
2003, v.16 no.4, pp.261-268






  • Downloaded
  • Viewed

Abstract

To investigate the matrix effect of sample for hydrogen analysis by inert gas fusion-thermal conductivity detection, calibration factor for the hydrogen analyser of the inert gas fusion-thermal conductivity detection method was measured with hydrogen standard materials in Ti, Zr-2.5Nb and by hydrogen gas dosing method. Also the hydrogen extraction efficiency for the different sample matrix, Ti and Zr-2.5Nb, was evaluated without adding tin flux. The calibration factor of the hydrogen analyser which was calibrated by hydrogen standard material in Zr-2.5Nb and Ti was 2~3% and 14% higher than that by hydrogen gas dosing method, respectively. Based on the results of calibration factor measurement, it could be concluded that the hydrogen extraction efficiency of the Ti matrix sample will be 12% lower than that of the Zr-2.5Nb. And according to the experimental results of hydrogen recovery test by no tin flux, the hydrogen recovery percentage of the Ti and Zr-2.5Nb matrix sample was about 70% but the recovery rate of Ti matrix sample was slightly lower than that of Zr-2.5Nb.

keywords
Ti, Zr-2.5Nb, hydrogen, matrix effect, Inert Gas Fusion-Thermal Conductivity Detection(IGF-TCD)