- P-ISSN 1225-0163
- E-ISSN 2288-8985
In this study, the separation of toxic dioxin 2,3,7,8-congeners by DB-5MS and SP-2331 GC columns which are widely used in HRGC/HRMS analysis was examined. Through the dioxin analysis of column performance check standard solution and fly ash sample, the isomer specific separation of 2,3,7,8-substituted dioxins from tetra to hexa-isomers on DB-5MS and SP-2331 columns were studied. The effect of I-TEQ value by these columns was also studied. The total concentrations of toxic dioxins for the column performance check standard solution were 508.4 ng/mL analyzed by DB-5MS and 515.8 ng/mL analyzed by SP-2331, respectively. The I-TEQ value obtained by both columns was shown to be almost equivalent for the column performance check standard solution and fly ash sample.
(dc.) EPA Method 1613"Tetra Through Octa-CDDs/Fs by Isotope Dilution HRGC/HRMS,