Comparison of Dioxin Analytical Result Between SP-2331 and DB-5MS Capillary columns in the HRGC/HRMS Anlaysis

Analytical Science and Technology / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
2004, v.17 no.2, pp.184-191





  • Downloaded
  • Viewed

Abstract

In this study, the separation of toxic dioxin 2,3,7,8-congeners by DB-5MS and SP-2331 GC columns which are widely used in HRGC/HRMS analysis was examined. Through the dioxin analysis of column performance check standard solution and fly ash sample, the isomer specific separation of 2,3,7,8-substituted dioxins from tetra to hexa-isomers on DB-5MS and SP-2331 columns were studied. The effect of I-TEQ value by these columns was also studied. The total concentrations of toxic dioxins for the column performance check standard solution were 508.4 ng/mL analyzed by DB-5MS and 515.8 ng/mL analyzed by SP-2331, respectively. The I-TEQ value obtained by both columns was shown to be almost equivalent for the column performance check standard solution and fly ash sample.

keywords
DB-5MS, Sp-2331, dioxin separation, fly ash, column performance check standard solution


Reference

1

(dc.) EPA Method 1613"Tetra Through Octa-CDDs/Fs by Isotope Dilution HRGC/HRMS,