HPLC/ESI/MS를 이용한 물 중의 알킬페놀에톡실레이트 분석

Determination of alkylphenol ethoxylate in water by high performance liquid chromatography/electrospray ionization/mass spectrometry

분석과학 / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
2004, v.17 no.3, pp.263-270
정봉철 (한국과학기술연구원)
이정애 (한국과학기술연구원)
박송자 (한국과학기술연구원)
  • 다운로드 수
  • 조회수

초록

알킬페놀에톡실레이트 (APEO)는 주로 비이온성 계면활성제로 쓰이고 있으며 세제, 액체연료 및 농약에도 이용이 되고 있다. APEO 자체는 독성물질로 분류되어 있지 않지만, 그 대사생성물질인 단사슬 APEO, 알킬페놀 및 카르복실 유도체등은 폐수 또는 음용수를 염소처리하는 동안 mutagenic ring halogenated derivative를 생성한다. 이들 생성물들은 estrogenic effect를 나타내기 때문에 APEO를 제한 또는 규제하고 있는 추세이다. 본 연구에서는 APEO의 대사체로서 단사슬 APEO인 4-nonylphenol-di-ethoxylate (NP2EO), 4-octylphenol-di-ethoxylate (OP2EO)를 분석하기 위해서 p-n- nonylphenol--di-ethoxylate-ring-13C6을 내부표준물질로 사용하여 HPLC/ESI/MS를 이용하여 분석하였다. 분석조건은 이온화 용매인 trifluoroacetic acid가 10 M이 되도록 각각 물과 메탄올에 첨가시켜 사용하였다. 물시료 1 L에 진한 황산을 이용하여 pH를 2이하로 조정한 후, 아세톤, 메탄올 및 물 (pH 2)로 활성화시킨 Sep-Pak C18에 loading 시킨 후 아세톤으로 용출하여 시료용액으로 하였다. 이 방법으로 농도범위가 20 500 ng/L 내에서 검량선의 직선성은 r=0.999 (OP2EO)와 0.990 (NP2EO)였으며, 검출한계는 OP2EO는 20 ng/L, NP2EO는 50 ng/L 이었다. 정확도 및 정밀도는 85.8 122.1% 및 8.2 18.8 %로 좋은 결과를 나타냈다. 이 방법은 환경시료로부터 미량의 APEO를 분석하는데 사용될 수 있고, APEO 오염실태 조사에 응용될 수 있을 것으로 사료된다.

keywords
alkylphenol ethoxylate, electrospray mass spectrometry

Abstract

A method is described for the analysis of short-chain alkylphenol ethoxylates (APEOs), 4-octylphenol-di-ethoxylate (OP2EO) and 4-nonylphenol-di-ethoxylate (NP2EO), in drinking water or wastewater using reversed phase high-performance liquid chromatography with electrospray ionization mass spectrometry. The solvent system was water and methanol containing 10 M trifluoroacetic acid as an ionization solvent. We acidified 1 L of water samples to less than pH 2 with concentrated H2SO4 and loaded onto Sep-Pak C18, and eluted with acetone. The calibration of OP2EO and NP2EO was performed for the concentration range from 20 to 500 ng/L and the correlation coefficients were 0.999 and 0.990, respectively. The limits of detection were 20 ng/L (OP2EO) and 50 ng/L (NP2EO) at a signal-to-noise ratio of 3. Accuracy and precision of this analytical method were 85.8∼122.1% and 8.2∼18.8%, respectively. The proposed method allowed a sensitive and rapid detection of OP2EO and NP2EO and it could be applied for monitoring of APEOs from environmental samples.

keywords
alkylphenol ethoxylate, electrospray mass spectrometry


참고문헌

1

Tanghe, T, (1998) Water Res.,

2

Ferguson, P. L. , (2000) Anal. Chem,

3

Reinhard, M, (1982) Environ. Sci. Technol.,

4

Sonnenschein, C, (1998) J. Steroid Biochem. Mol. Biol.,

5

Jobling, S, (1996) Environ. Tixicol. Chem.,

6

Shang, D. Y, (1999) J. Chromatogr. A,

7

Petrovic, M, (2001) Anal. Chem.,

8

Moreno-Garrido, I, (2001) Fresenius J. Anal. Chem.,

9

La Guardia, M. J, (2001) Environ. Sci. Technol.,

10

Snyder, S. A, (2001) Environ. Toxicol. Chem.,

11

Ahel, M, (1985) Anal. Chem.,

12

Boyd-Boland, A. A, (1996) Anal. Chem.,

13

Marcomini, A, (1990) Marine Chem,

14

Hoffman, B. J, (2002) J. Chromatogr. Sci.,

15

Shao, B, (2002) J. Chromatogr. A,

16

Cohen, A, (2001) J. Chromatogr. A,

17

Petrovic, M, (2001) J. Mass Spectrom.,

18

Reemtsma, T, (2001) Trands in Anal. Chem.,

19

Crescenz, C, (1995) Anal. Chem.,

상단으로 이동

분석과학