Direct determination of gadolinium in urania-gadolinia nuclear fuels by inductively coupled plasma atomic emission spectrometry

Analytical Science and Technology / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
2007, v.20 no.2, pp.131-137





  • Downloaded
  • Viewed

Abstract

The urania-gadolinia fuels were dissolved with nitric acid. The analytical conditions of ICP-AES for the direct determinations of gadolinium in the uranium matrices without separation process were investigated. Based on the effect of uranium on gadolinium intensity, the best wavelength for gadolinium was 336.223 nm. The relative deviation of two methods, direct and indirect measurements with anion exchange chromatography, was less than 5 %. Therefore it was possible for this procedure directly to measure 5~10 wt.% of gadolinium in urania-gadolinia fuels without separation by ICP-AES.

keywords
urania-gadolinia fuel, direct determination, gadolinium, ICP-AES


Reference

1

R. A. Storer, (1987) 1987 Annual Book of ASTM Standards,

2

G. R. Harrison, (1969) Massachusetts Institute of TechnologyWavelength Tables, M.I.T. Press

3

H. O. Haug, (1982) Gmelin Handbook of Inorgnaic Chemistry, Springer-Verlag