Abstract

In our study, the accuracy for measurement of seventeen 2,3,7,8-substituted PCDDs/PCDFs in certified reference material (CRM) which is the sample of homogeneous sediment matrix taken from an area known to have significant chemical contamination, particularly PCBs (polychlorinated biphenyls), was evaluated. Though the methodology in this study followed the official method of unintentionally produced persistent organic pollutants (UPOPs) announced by the Ministry of Environment of the Republic of Korea in 2007, there were slight changes using additional purification step by activated carbon column because the interferences of sample were not sufficiently removed when only multi-silica column and alumina column have been used for purification. The |En| number proposed by the Korea Research Institute of Standards and Science was used for a valuation basis of the accuracy. The |En| numbers of seventeen 2,3,7,8-substituted PCDDs/PCDFs have been indicated as 1 and below, they were decided “Pass” in this test, when DB-5MS column and SP-2331 column were used together. Because 1,2,3,7,8-PeCDD and #169-HxCB were not separated on DB-5MS column, the ions of 1,2,3,7,8-PeCDD were selected at M/M+2 instead of M+2/M+4 suggested by EPA 1613. It is possible to distinguish them in HRGC/HRMS analysis.

keywords
PCDDs/PCDFs, CRM, accuracy, DB-5MS, SP-2331


Reference

1

1. E. J. Kim, J. E. Oh and Y. S. Chang, The Science of the Total Environment, 311, 177-189(2003).

2

2. J. E. Oh, Y. S. Chang, E. J. Kim and D. W. LEE, Atmosp. Environ., 36, 5109-5117(2002).

3

3. Y. H. Yang, Y. S. Chang, B. H. Kim, D. C. Shin and M. G. Ikonomou, Chemosphere, 47, 1087-1095(2002).

4

4. K. S. Kim, S. K. Shin, K. S. Kim, B. J. Song and J. G. Kim, Environ. Internat., 34, 202-209(2008).

5

5. S. J. Kim, J. G. Kim, T. W. Jeon, S. K. Shin and K. S. Kim, Bull. Environ. Contam. Toxicol. 83(3), 435-439 (2009).

6

6. 환경부 고시 제 2007-165호 : 비의도적 잔류성유기 오염물질 공정시험방법.

7

7. B. W. Yu, Y. H. Moon, M. K. Kim, J. D. Kyoung, Y. S. Chang, Anal. Sci. Tech., 17(2), 153-162(2004).

8

8. 허귀석박사 초청강연, 한국표준과학원, 2008 춘계 GS/LG/LS 분석기술교류회.

9

9. Certificate of Analysis, Heavily Contaminated Sediment Reference Material (EDF-5184), CIL.

10

10. 환경부 고시 제 2007-165호 :비의도적 잔류성유기오염물질 공정시험방법 ES 10428.1 (토양/퇴적물 시료중 다이옥신 및 퓨란류 시험방법 - HRGC/HRMS).

11

11. JIS K0312, 2006 Method for determination of tetrathrough octa-chlorodibenzo-p-dioxins, tetra-through octachlorodibenzofurans and coplanar polychlorobiphenyls in industrial water and waste water.

12

12. EPA, Method 1613, Tetra through octachlorinated dioxins and furans by isotopic dilution HRGC-HRMS, Washington DC, 1994.

13

13. ‘Guide to the expression of uncertainty in measurement’, ISO, Geneva, Switzerland, 1993.

14

14. 다이옥신 분석 불확도 추정보고서(소각재), 경기도 보건환경연구원.

15

15. S. J. Mun, B. H. Kim, J. C. Woo, Y. S. Chang, Anal. Sci. Tech., 16(6), 475-482(2003).

16

16. Y. G. Ahn, J. H. Shin, S. Y. Yoo, J. H. Khim, J. K. Hong, Anal. Sci. Tech., 18(1), 13-22 (2005).

17

17. J. J. Ryan, H. B. S. Conacher, L. G. Panopio, B. P. Y. Lau and J. A. Hardy, J. Chromatogr., 541, 131-183(1991).