Article Detail

Home > Article Detail
  • P-ISSN 1225-0163
  • E-ISSN 2288-8985

A polymer pH-Selectrode Based on Tribenzylamine as Neutral Carrier

Analytical Science and Technology / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
1995, v.8 no.1, pp.63-68
Park, Myon-Young
Chung, Koo-Chun
Cho, Dong-Hoe
Lee, Kyeong-Jae
Jeong, Seong-Suk
Park, Sun-Young
Kim, Tae-Hun
  • Downloaded
  • Viewed

Abstract

For the preparation of pH-selectrode, tribenzylamine, polyvinylchloride, dioctylphthalate, sodium tetraphenylborate and tetrahydrofuran were mixed with 0.02, 0.62, 1.34, 0.02g and 10ml respectively, and added 1g of acetylene black, graphite, silicon carbide or tungsten carbide respectively to improve electric conductivity. The selectrodes of seven kinds were shown linear to hydrogen ion in the range of pH 2 and 9. The best electric conductor for preparation of pH-selectrode based on tribenzylamine as neutral carrier was acetylene black and responded potential of the selectrode to hydrogen ion was shown the values near to theoretical Nernstian slope at <TEX>$20^{\circ}C$</TEX>. The interfering effects of the selectrode on hydrogen ion in the presence of alkali and alkaline earth metal ions were shown the better results with less error than glass electrode. The reproducibility and stability were good for use as a selectrode, especially in the presence of fluoride ion.