Dynamic Range in the Simultaneous Analysis of Pesticide Priority Pollutants on Mass Spectrometer

Analytical Science and Technology / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
1996, v.9 no.3, pp.320-324
Lee, Seung-Seok
Lee, Sueg-Geun
  • Downloaded
  • Viewed

Abstract

The dynamic range of mass selected detector, which is very popular recently because of the simplicity and cheap price, was determined by using priority pollutants. The applied priority pollutants were four pesticides, 4,4'-DDE and 4,4'-DDD, which show the sharp peaks, and <TEX>${\alpha}$</TEX>-endosulfan and <TEX>${\beta}$</TEX>-endosulfan, which show the broad peaks in SIM Chromatogram. The dynamic range of the 4,4'-DDE and the 4,4'-DDD was <TEX>$0.1{\sim}3000{\mu}g/ml$</TEX>. The broad peaks of the <TEX>${\alpha}$</TEX>-endosulfan and <TEX>${\beta}$</TEX>-endosulfan resulted in <TEX>$0.25{\sim}3000{\mu}g/ml$</TEX> of dynamic range.