Development of a Multi - channel Detector for Capillary Electrophoresis System

Analytical Science and Technology / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
1998, v.11 no.2, pp.96-104
Hong, Seung Guk
Kim, Hai-Dong
  • Downloaded
  • Viewed

Abstract

A photodiode array multichannel detector system for capillary electrophoresis was developed. The photodiode array detector for capillary electrophoresis (CE-PDA) has 1024 photodetectors and can analyze sample by measuring UV/VIS absorption spectrum in 275~675 nm wavelength range. The CE-PDA instrument can get a spectrum in 30 ms during sample separation and can be programmed by a PC to control various experimental conditions required for sample analysis. The performance of the multichannel CE-PDA instrument was tested using L-ascorbic acid and alizarin yellow GG mixture. The reproducibility test of the CE-PDA system showed 5.6% RSD.

keywords
CE, photodiode array, multi-channel detector