Determination of Ni, Cr, Mo in Low Alloy Steel Reference Materials by Isotope Dilution Inductively Coupled Plasma Mass Spectrometry

Analytical Science and Technology / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
2003, v.16 no.1, pp.82-89




  • Downloaded
  • Viewed

Abstract

Isotope dilution mass spectrometry (IDMS) was applied to the determination of Ni, Cr, Mo in low alloy steel reference materials. The Mo isotope ratio measurement was performed by dynamic reaction cell inductively coupled plasma mass spectrometry (DRC-ICP/MS) using ammonia as a reaction cell gas. In the case of Ni and Cr measurement, all data were obtained at medium resolution mode (m/<TEX>${\Delta}m

keywords
IDMS, DRC-ICP/MS, HR-ICP/MS, Ni, Cr, Mo, low alloy steel, NIST SRM