    Study of improving precision and accuracy by using an internal standard in post column isotope dilution method for HPLC-ICP/MS

    Analytical Science and Technology / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
    2014, v.27 no.3, pp.140-146
    https://doi.org/10.5806/AST.2014.27.3.140



    • Downloaded
    • Viewed

    Abstract

    An internal standard was used in PCID (post column isotope dilution) to improve the accuracy andprecision in quantification of various chemical species. The error occurring in the column was the largest inHPLC-ICP/MS (high performance liquid chromatography-inductively coupled plasma/mass spectrometry) whenPCID and other traditional quantification methods were compared with each other. Internal standard was effectivein correcting the loss of sample in the column to improve accuracy and precision. When applied to SeMet, usingMeSecys or Se4+as an internal standard, relative errors were reduced from 31% and 13% to less than 1%, whilestandard deviations were reduced from 5.1% and 6.9% to 1.5% and 0.2%, respectively. Positive aspects of usingan internal standard in PCID were compared with other quantitative techniques and discussed in detail.

    keywords
    internal standard, post column isotope dilution, species unspecific ID, HPLC-ICP/MS


    Reference

    1

    1. C. Swart, O. Rienitz and D. Schiel, Talanta, 83, 1544-1551 (2011).

    2

    2. K. G. Heumann, Int. J. Mass. Spectrom. Ion Process., 118/119, 575-592 (1992).

    3

    3. L. Rottmann, K.G. Heumann, Fresenius J. Anal. Chem., 350, 221-228, (1994).

    4

    4. M. Fernanda Giné and A. Paula Packer, J. Braz. Chem. Soc., 21, 575-589 (2010).

    5

    5. H. A. Meyer, B. Hollenbach, C. Stephan, T. Endermann, N. G. Morgenthaler, H. Cammann, J. Köhrle, K. Jung and L. Schomburg, Cancer Epidemiol Biomarker Prev., 18(9), 2386-2390 (2009).

    6

    6. P. Rodrigez-Gonzalez, J. M. Marchante-Gayon, J. I. Garcia-Alonso and A. Sanz-Medel, Spectrochimica Acta, 60B, 151-207 (2005).

    7

    7. H. G. Infante, K. V. Campenhout, R. Blust and F. C. Adams, J. Chromatogr. A, 184, 1121-1129 (2006).

    8

    8. A. Rodríuez-Cea, M. R. Fernádez, E. B. Gonzalez, B. A. Fernádez and A. Sanz-Medel, J. Anal. At. Spectrom., 18, 1357-1362 (2003).

    9

    9. K. V. Campenhout, H. G. Infante, G. Goemans, C. Belpaire, F. C. Adams, R. Blust and L. Bervoets, Sci. Total Environ., 194, 379-386 (2008).

    10

    10. A. Rodríuez-Cea, M. R. Fernádez, J. I. Garcí-Alonso and A. Sanz-Medel, J. Anal. At. Spectrom., 21, 270-278 (2006).

    11

    11. H. G. Infante, K. V. Campenhout, D. Schaumlöfel, R. Blust and F. C. Adams, Analyst, 128, 651-660 (2003).

    12

    12. J. Jeong, J. Lee and Y. Pak, Bull. Kor. Chem. Soc., 34, 3817-3824 (2013).

    13

    13. K. Polec-Pawlak, D. Schaumlöfel, J. Szpunar, A. Prange and R. Lobinski, J. Anal. At. Spectrom., 17, 908-913(2002).

    14

    14. D. Schaumlöfel, A. Prange, G. Marx, K. G. Heumann and P. Bräter, Anal. Bioanal. Chem., 55, 372-378 (2002).

    15

    15. H. Jang, H. Min, J. Lee and Y. Pak, Anal Sci and Technol., 26, 182-189 (2013).

    16

    16. E. Kim, M. Joo, H. Kwon and Y. Pak, Anal Sci and Technol., 26, 307-314, (2013).

    17

    17. M. Kotrebai, J. Tyson, E. Block and P. C. Uden, J. of Chromatography A, 866 51-63, (2000).

    18

    18. O. Rienitz A. Pramann and D. Schie, International Journal of Mass Spectrometry, 289, 47-53 (2010).

    상단으로 이동

    Analytical Science and Technology