Article Detail

Home > Article Detail
  • P-ISSN 1225-0163
  • E-ISSN 2288-8985

Study on the Analytical Performance of Double Membrane Desolvator(DMD) with Perfluoroalkoxy(PFA) and Microconcentric Nebulizer(MCN) in Inductively Coupled Plasma Atomic Emission Spectrometry

Analytical Science and Technology / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
2001, v.14 no.1, pp.21-27
Sul, Choon-Kuen
Lim, H.B.
  • Downloaded
  • Viewed

Abstract

In this work, analytical performance of a laboratory built double membrane desolvator (DMD) was studies using perfluoroalkoxy (PFA) nebulizer and microconcentirc nebulizer (MCN) in inductively coupled plasma atomic emission spectrometry. Compared with MCN, PFA nebulizer coupled to DMD showed similar analytical sensitivity for aqueous solution and better sensitivity for isopropyl alcohol. Since PFA resisted various acids, HCI, <TEX>$H_2SO_4$</TEX> and HF solution were analyzed with less than 2% RSD. Rinse-out time for the signal reduction to 1% was obtained to be 35 s for PFA but about 45 s for MCN.

keywords
ICP-AES, membrane desolvator, PFA nebulizer, MCN


상단으로 이동