Optimization of Response Characteristics of pH-ISFET Glucose Sensor

Analytical Science and Technology / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
1992, v.5 no.2, pp.177-184
Lee, Heung Lark
Yang, Seung Tae
Jung, Doog Sook
Kim, Chang Soo
Sohn, Byung Ki
  • Downloaded
  • Viewed

Abstract

A preparation method and response characteristics of a glucose sensor which consisted of pH-ISFET and glucose oxidase-immobilized membrane were investigated. The pH-ISFET glucose sensor was fabricated by immbilizing bovine serum albumin and glucose oxidase with glutaraldehyde on gate of the pH-ISFET. Effects of pH and concentration of working buffer and enzyme load on the pontentiometric response of the pH-ISFET glucose sensor were examined. Response characteristics for the determination of glucose in synthetic physiological saline solution(pH 7.4) were as follows. That is the concentration range of linear response, slope of linear response(sensitivity), and response time were <TEX>$1.0{\times}10^{-4}{\sim}6.0{\times}10^{-3}M $</TEX>, 4.1 mV/decade, and 12~15 min., respectively.

keywords
Glucose sensor pH-ISFET