Characterization of Atomic Emission Detector for Gas Chromatography Using Cylindrical Microwave Cavity

Analytical Science and Technology / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
1992, v.5 no.3, pp.263-268
Park, Young-Joo
Yoo, Hee-Soo
  • Downloaded
  • Viewed

Abstract

A plasma source with cylindrical microwave cavity was used as atomic emission detector for gas chromatography. Detection limits of several elements were determined for this system. Detection limits for bromine and sulfur were 0.46 pg/s and 0.51 ng/s, respectively. The plasma was stable at the range of flow rate of 10 to 20mL/min.

keywords
GC/MIP Detector, Atomic Emission Detector, Detection Limit