A Study on the Precise End-Point Detection in Titration by Using the Phase Angle Measurements

Analytical Science and Technology / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
1999, v.12 no.4, pp.290-298
Park, Byung-Bin
Shin, Ho-Sang
Lee, Han-Hyoung
  • Downloaded
  • Viewed

Abstract

A study on the application of impedance phase angle for redox titration, acid-base titration, chelate titration and precipitation titration has been carried out. A constant alternating current was passed between two platinum electrodes. One of them was a polarizable micro-electrode of <TEX>$0.1cm^2$</TEX> or <TEX>$0.026cm^2$</TEX> surface area and the other a non-polarizable large electrode of <TEX>$1cm^2$</TEX> surface area dipped in the solution to be titrated. The impedance and the phase angle of the titration cell were measured with lock-in amplifier to obtain well behaved titration curve respectively. In titration of oxalic acid vs. potassium permanganate, the end-point was obtained successfully from the phase angle titration curve. In this experiment, the concentration of 0.0005 M to 0.05 M, the current of <TEX>$50{\mu}A$</TEX> and the frequency of near 50 Hz were used. In titration of phosphoric acid vs. sodium hydroxide, the first end-point was obtained successfully on the optimum experimental condition of 0.001 M concentration, <TEX>$50{\mu}A$</TEX> current and 25~97 Hz frequency. However, the end-point in titration of cupric sulfate vs. disodium-EDTA couldn't be obtained clearly. The end-point was obtained with the out-of-phase impedance curve on the experimental condition of 0.01 M concentration, <TEX>$100{\mu}A$</TEX> current, 5~35 Hz frequency range. In titration of sodium chloride vs. silver nitrate, the end-point was obtained successfully on the experimental condition of 0.1 M concentration, <TEX>$100{\mu}A$</TEX> current and 5~47 Hz frequency range. This study showed that the impedance phase angle was applicable for the detection of the end-points in redox titration curve, acid-base titration curve, chelate titration curve and precipitation titration curve.

keywords
impedance phase angle, end-point, titration curve


상단으로 이동

Analytical Science and Technology