Construction and Evaluation of Cryogenic Preconcentration System for Analysis of VOC with ppb Level in the Air

Analytical Science and Technology / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
1999, v.12 no.4, pp.332-340
Jeon, Sun Joo
Kim, Byung Joo
Kim, Jin Seog
Heo, Gwi Suk
  • Downloaded
  • Viewed

Abstract

Homemade cryogenic preconcentration system is constructed to preconcentrate the VOCs in the ambient air collected in canister. Homemade preconcentrator equipped with GC-MS was used for analysis of air sample containing VOCs at ppb level identified in EPA TO-14 method. The system was evaluated for analytical performance by comparison with the results of adsorption tube method widely used in the past, and shows good agreement for concentrations of benzene, toluene, ethylbenzene, and xylenes.

keywords
cryogenic preconcentration, canister, VOC, GC-MS