- P-ISSN 1225-0163
- E-ISSN 2288-8985
The separation and determination of inorganic cations by a capillary zone electrophoresis was studied. Cations were separated by the capillary column and detected by indirect UV method. The running buffer of 0.03 M creatinine and 0.02 M-hydroxy isobutylic acid solution (pH 4.8) was used to improve the mobility of ions in the column. A potential of 18 kV was applied at anode. A complete separation of cations ranged within 15 min and detection limits were between 0.1 and 1.0 ppm. The applicability of the method for real samples was demonstrated.