Article Detail

Home > Article Detail
  • P-ISSN 1225-0163
  • E-ISSN 2288-8985

Article Contents

    Development and validation of a selective and sensitive LC-MS/MS method for determination of misoprostol acid in human plasma: Application to bioequivalence study

    Analytical Science and Technology / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
    2015, v.28 no.1, pp.17-25
    https://doi.org/10.5806/AST.2015.28.1.17
    Yong Sun Park (Bio-Medieng Co. Ltd)
    Keun Nam Kim (Bio-Medieng Co. Ltd)
    Ye Na Kim (Bio-Medieng Co. Ltd)
    Jung Hwan Kim (Bio-Medieng Co. Ltd)
    Jin Young Kim (Bio-Medieng Co. Ltd)
    Soo Ji Sim (Bio-Medieng Co. Ltd)
    Heon Soo Lee (Bio-Medieng Co. Ltd)
    • Downloaded
    • Viewed

    Abstract

    A rapid, sensitive and specific method was developed and validated using electrospray ionization(ESI) tandem mass spectrometry (LC-MS/MS) for determination of misoprostol acid in human plasma. Misoprostolacid-d5 was used as in internal standard (IS). The analyte and IS were extracted by simple one step solid phaseextraction (SPE). Linearity in plasma was obtained over the concentration range 10~3000 pg/mL and lowerlimit of quantification (LLOQ) was identifiable and reproducible at 10 pg/mL. The intra- and inter-day precisionvalues were below 9% and the accuracy was ranged from 93.81% to 102.02% at all four quality control samples. The method was has been successfully applied for routine assay to support pharmacokinetic study of misoprostolacid in human plasma after an oral administration of 0.4 mg misoprostol.

    keywords
    misoprostol acid, alsoben, human plasma, LC-MS/MS, pharmacokinetics


    Reference

    1

    1. P. W. Collins, Med. Res. Rev., 110, 149-172 (1990).

    2

    2. R. D. Newman, N. Citlin, E. J. Lacayo, A. V. Safdi, E. J. Ramsey, S. L. Engel, A. Rubin, C. H. Nilsen and E. A. Swabb, Am. J. Med., 83(1A), 27-31 (1987).

    3

    3. A. Robert, Gastroenterology, 77, 761-767 (1979).

    4

    4. Metin Gülmezoglu, José Villar, Nguyen Thi Nhu Ngoc, Gilda Piaggio, Guillermo Carroli and Lekan Adetoro, et al, Lancet, 358, 689-695 (2001).

    5

    5. G. Schoenhard, J. Oppermann and F. E. Kohn, Dig. Dis. Sci., 30(11), 126S-128S (1985).

    6

    6. A. Karim, L. F. Rozek, M. E. Smith and K. G. Kowalski, J. Clin. Pharmacol, 29, 439-443 (1989).

    7

    7. E. F. Foote, D. R. Lee, A. Karim, W. F. Keane and C. E. Halstenson, J. Clin. Pharmacol, 35(4), 384-9 (1995).

    8

    8. M. Zieman, S. K. Fong, N. L. Benowitz, D. Banskter, P. D. Darney, Obstet Gynecol, 90(1), 88-92 (1997).

    9

    9. O. S. Tang, H. Schweer, H. W. Seyberth, S. W. Lee and P. C. Ho, Hum Reprod, 17(2), 332-336 (2002).

    10

    10. B. Watzer, H. W. Seyberth and H. Schweer, J. Mass Spectrom, 37(9), 927-33 (2002).

    11

    11. H. Abdel-Aleem, J. Villar, A. M. Gülmezoglu, S. A. Mostafa, A. A. Youssef, M. Shokry and B. Watzer, Eur J. Obstet Gynecol Reprod Biol., 108, 25 (2003).

    12

    12. K. G. Danielsson, L. Marions, A. Rodriguez, B. W. Spur, P. Y. Wong and M. Bygdeman, Obstet Gynecol, 93(2), 275-280 (1999).

    13

    13. K. L. Andolina, J. E. Tolosa, J. M. Monzo, N. S. Roberts and S. Daly, J. Matern. Fetal Neonatal Med., 14(4), 229-232 (2003).

    14

    14. Y. Zou, X. Chen, B. Song and D. Zhong, J. Chromatography B., 852, 122-127 (2007).

    15

    15. C. Fiala, A. Aronsson, F. Granath, O. Stephansson, H. W. Seyberth, B. Watzer and K. Gemzell-Danielsson, Hum Reprod., 20(12), 3414-3418 (2005).