    Direct Analysis of Aerosol Particles by Atomic Emission and Mass Spectrometry

    Analytical Science and Technology / Analytical Science and Technology, (P)1225-0163; (E)2288-8985
    1995, v.8 no.4, pp.411-418
    Kawaguchi, Hiroshi
    Nomizu, Tsutomu
    Tanaka, Tomokazu
    Kaneco, Satoshi
    • 0Downloaded
    • 290Viewed

    Abstract

    A method for the direct determination of elemental content in each of aerosol particles by inductively coupled plasma atomic emission (ICP-AES) or mass spectrometry (ICP-MS) is described. This method is based upon the introduction of diluted aerosol into an ICP and the measurement of either the flash emission intensities of an atomic spectral line or ion intensities. A pulse-height analyzer is used for the measurement of the distribution of the elemental content. In order to calibrate the measuring system, monodisperse aerosols are used. The potentials of the method are shown by demonstrating the copper emission signals from the aerosols generated at a small electric switch, a study of the relation between the decreasing rate of particle number density and particle size, and measurements of calcium contents in the individual biological cells.

    keywords
    Inductively coupled plasma, Atomic emission spectrometry, Mass spectrometry, Aerosol particles, Monodisperse aerosol